| Name: | DCL4.2 RGbast online residual chlorine electrode | 
|---|---|
| Measuring range: | 0.005~0.50/2.00ppm 0.05~5.00/10.00/20.00ppm 0.5~100.0/200.0ppm | 
| Working temperature: | 0~45℃ | 
| Working pressure: | 1.0bar | 
| Output: | 4-20mA、Modbus485、0~±2V | 
| Measured medium | Free chlorine (depending on pH) | 
| Measuring principle | Coated two-electrode current method | 
| Measuring range | 0.005~0.50/2.00ppm 0.05~5.00/10.00/20.00ppm 0.5~100.0/200.0ppm | 
| pH range | 6.0~8.0 | 
| Working temperature | 0~45℃ | 
| Flow speed | 15~30L/H | 
| Working pressure | 1.0bar | 
| Zero calibration | Not needed | 
| Slope calibration | According to DPD-1 method, calibrate on the device | 
| Drift | <-1%/month | 
| Response time | It takes 30 seconds to reach 90% of the measured value | 
| Polarization time | 1 hour | 
| Electrical connection | 4-pole plug, 5-pole M12 connector, 2-pole terminal | 
| Power supply | DC 12~30V, ±5~±15V | 
| Signal output | 2-wire 4-20mA, Modbus485, 0~±2V | 
| Dimensions | Diameter: 25mm Length: 175~220mm | 
| Material | Semi-permeable membrane, PVC-U, ABS | 
| Maintenance interval | Membrane cap: replace once a year, electrolyte: replace 3-6 months | 
| Storage | Electrode rod and membrane cap: 5~40℃ dry environment Electrolyte: 5~35℃ in a cool and dark place | 
| CE | EMC, RoHS | 
Non-surfactant-free drinking water, medical water, public service water and water quality monitoring at the end of the pipe network
Non-surfactant-free drinking water, medical water, public service water and water quality monitoring at the end of the pipe network
Non-surfactant-free drinking water, medical water, public service water and water quality monitoring at the end of the pipe network
 
                        